Eligibility of Subject studied -
The test subject was undefined and of arbitrary choice.
The test cases studied are 153, held at the Fortis hospital in Dehradun.
The test subject consent was taken prior to taking their test via SPANDAN ECG.
The objective of the study
To check whether the Spandan detects the Abnormality in Lead 2.
Eligibility of Spandan to perform with respect to 12 lead Gold standard ECG.
Validating the AHA standards to evaluate Abnormality detectionby Specificity and Sensitivity.
Introduction
Abnormality detection is the important feature of Spandan ECG, Hence its validation study was conducted to
validate the specificity and sensitivity of results produced by the Spandan ECG with respect to 12 lead Gold
standard. AHA describes the specificity and sensitivity as the correct method to evaluate the accuracy of the
results provided by any smartphone based ECG machine.
The Spandan ECG and results were evaluated and tested with respect to the 12 lead gold standard ECG. The machine
that is used here for validation of the results is GEMAC2000, which was used for testing in the health camp
organized by Fortis Hospital and Sunfox Technologies in Dehradun.
Methodology
The evaluation study performed was under the protocol set by the Spandan technical team, which was unbiased and
partial and involve the involvement of Cardiologist and a physician. The Interpretation of Abnormality by
Spandan was taken as the characteristic attribute for the validation study. The study was aimed to achieve the
minimum deviation in the lead 2 of Spandan ECG and GEMAC2000 Gold standard.
The tests were performed in the
subjects with consideration of the following:
The subject consent must be taken for the test.
There must be not the difference of five minutes between the test taken by Spandan ECG and GEMAC2000.
The cases with baseline wander are excluded
Cases with motion artifacts are excluded from results.
Cases with wrinkled skin are excluded.
People with chest pain are included in the study.
Normal cases are also included.
The test must be taken by 12 lead Gold standard at first.
The following are the combinations used as True test and False test
Spandan/12 leadGoldstandard |
Abnormal/NSR |
Abnormal/DEFECT |
Normal/NSR |
Normal/Defect |
Abnormal/NSR |
true |
true |
true |
false |
Abnormal/Defect |
true |
true |
true |
false |
Normal/NSR |
true |
true |
false |
true |
Normal/Defect |
false |
true |
true |
true |
Here are the steps to be followed by the researcher for the validation as shown in the flow chart:-
Results and Discussions
The test performed will be evaluated with a human error of 5% will be taken into consideration.
The Deviation in the 153 subjects is shown in the figure below. Through which there is an accuracy of 98.5
percent in interpreting the Abnormality and normality in the cases.
Total cases |
True Detection |
False detection |
153 |
150 |
3 |
Accuracy |
98.5% (With Human Error) |
|
The figure below shows the data of the tested subject’s normality. The ability to detect the normal and
abnormality Via. chest lead 2 is near to 98.5%.The results were comprised of false positives, which was the
vantage point for our work in designing the algorithms for Spandan ECG
As per the AHA, there isnocomputerprogram that can provide the correct interpretation as a skillful physician,
hence the challenge was to bring the accuracy as near to the 12 lead Gold standard ECG. The patients were
divided into two groups.
Group 1 – Patients with Known CVD or any Abnormality
Group 2 – Patients with known Normality
This consideration was taken due to the in-depth analysis of the accuracy of Spandan in delivering better and
accurate results as the 12 lead Gold standard.
12 Lead Gold standard |
Abnormal Cases |
Normal Cases |
Borderline cases |
Number of cases |
91 |
54 |
8 |
The accuracy is calculated by using the following formula:-
% Accuracy = (Result via.12 lead Goldstandard-Result via Spandan)/(Results via 12 lead
Goldstandard) x100
The data with the following combinations of results were neglected and reclassified as normal or abnormal:-
Borderline ECG in Spandan and Borderline ECG in 12 lead Gold standard is True result.
Borderline ECG in Spandan and Abnormal ECG in 12 lead Gold standard is a true result.
Borderline ECG in Spandan and Normal ECG in 12 lead Gold standard is the true result.
Abnormal ECG in Spandan and Borderline ECG in 12 lead Gold standard is a true result.
Normal ECG in Spandan and Borderline ECG in 12 lead Gold standard is the true result.
Hence For Normality detection in Group 2 via. Spandan ECG refers to the following table, which includes cases
with 54 normal subjects,and 4 cases of Borderline subjects as interpreted by 12 lead Gold standard.
Case detected |
True Detection |
False Detection |
Number of cases |
57 |
1 |
Accuracy |
98.3% |
|
For Abnormality, detection is Group 1 via. Spandan ECG refers to the following table, which includes the cases
of 91 abnormal subjects, and 2 Borderline subjects as interpreted by 12 lead gold standard.
Case detected |
True Detection |
False Detection |
Number of cases |
90 |
3 |
Accuracy |
96.77% |
|
The Two Borderline ECG cases were 100% accurate in detection and Interpretation. Hence, the accuracy of the test
overall was 98.5 %.
Conclusion and Scope
The Spandan ECG is 98.5% accurate as compared to the 12 lead Gold standard (GEMAC2000) in Interpreting the
similar results which are under permissible limits of +-5% error by medical standards.
The Spandan ECG meets the Glasgow measurement accuracy standards for QT Interval.
Spandan is suitable for detection of heart abnormalities as there are No False positives.
Can be used by Cardiac patients and Elderly people for monitoring and prevent detection.
Recommended for daily check-ups at home.
Recommended for Small clinics and Skilful Physicians.